Source: http://www.doksinet Prof. dr Davor Eterović, editor Authors: D. Eterović, D Kovačić, V Marković, A Punda Laboratory exercises: Medical physics and biophysics/2012-13 Department of Medical Physics and Biophysics University of Split – School of Medicine Source: http://www.doksinet Original title: Upute za vježbe – Medicinska fizika i biofizika Katedra za medicinsku fiziku i biofiziku, Sveučilište u Splitu-Medicinski fakultet Translated by: 1. M Raguž (exercises 1,2,5,8 and 9) 2. D Eterović (exercises 3,4 and 10) 3. D Kovačić (exercises 6 and 7) Source: http://www.doksinet Laboratory exercises: Medical physics and biophysics 2012/13 Students are obliged to study the following instructions for exercises and to consult the texts of the corresponding lectures. Exercise Place E1: Radioactivity and radiation protection Dept. of Nucl Med E2: Absorbed radiation dose Dept. of Nucl Med E3: Measurements in nuclear medicine E4: Radiogram contrasts Dept. of

Nucl Med BSB: Laboratory E5: Echogram resolution Dept. of Oncology E6: Neural signal BSB: Laboratory E7: Audiometry BSB: Laboratory E8: Optical bench BSB: Laboratory E9: Viscosity BSB: Laboratory E10: Hemodynamic BSB: Laboratory 1. Department of Nuclear Medicine; University Hospital Split; location Firule, Spinčićeva 1 2. Department of Oncology; University Hospital Split; location Firule, Spinčićeva 1 3. BSB: Laboratory; Basic Science Building, Medical Faculty Split; location Križine, Šoltanska 2 Source: http://www.doksinet Exercise 1: Radioactivity and protection of radiation Radioactivity: Process in which energetically unstable atomic nuclei spontaneously changes (decay) because of the transition to a lower energy state. Law of radioactive decay: N = N0 e -λt N0 – the initial number of radioactive nuclei N – the number of radioactive nuclei remaining at time t λ – the decay constant t – time e = 2.72 Half-life is the time taken for half the

radioactive nuclei to decay T1/2 = 0.693 /  Thickness of the absorber When radiation passes through matter, the intensity of radiation decays because of the partial or total absorption of radiation in above mentioned matter. The portion of absorbed radiation depends on: type of radiation, energy of radiation and characteristics of the absorber. Source: http://www.doksinet N = N0 e –(/)d N0 – the initial number of radioactive nuclei N – the number of radioactive nuclei unabsorbed or undeflected from the straight beam after passing through the absorber of thickness d / – mass attenuation coefficient d – thickness of absorber Determination of the half-value thickness of the absorber by graphical method: Instructions for exercise: Geiger-Müller (G-M) tube is an instrument widely used for detection of radiation. First, we need to measure basic (background) radiation during period of one minute. Background radiation is the amount of radiation present in the

room regardless source of radiation. It is necessary to subtract background radiation from each measurement of radiation from available sources. Why?  Adjust radioactive material among lead blocks. Place the probe connected to the G-M tube above it and measure the number of beats for one minute. Subtract background radiation in order to get N0.  Place the absorber of known thickness between the source of radiation and the G-M counter and repeat measurement in above described way.  Repeat measurement a few times for different thicknesses of the absorber and draw N-d dependence on millimetric paper.  Determine d/2 from the graph. Source: http://www.doksinet Protection from radiation 1. Principles of protection: - Source of radiation (type of radiation, dose) - Time - Distance (I~1/R2) - Shields 2. Radiation dosimetry: - Personal dosimeters (ionization chamber, radiographic film, thermoluminiscent dosimeter) Demonstrate influence of each particular principle of protection

at decrease of exposure using personal dosimeter (ionization chamber). Measurement of the relative gamma constant (Γ) Read lesson Dosimetry in the chapter Ionizing radiation, contained in the material entitled Biophysics lectures 2011-2 /Diagnostics (PDF file at the Web site). You have an access to two different sources of ionization radiation: 131I and 99mTc. 1. Measure their activities using G-M tube 2. Measure rates of exposures at the same ‘spot’ in the space (at the distance R from each particular source of radiation) using personal dosimeter. 3. Calculate gamma constant of 131I based on performed measurements and using the next equation: Rate of exposure =ΓA/R2 Gamma constant of radioactive isotope 99mTc is equal 0.14 Cm2/kgBqs x 1014 Source: http://www.doksinet Exercise 2: Absorbed dose The subject of exercise and what you should do This exercise shows the method of planning the therapy of hyperfunctioning thyroid gland with exposure to ionizing radiation from the beta

emitter I-131, given orally in form of potassium or sodium iodide. Upon entering the circulation, the radioactive iodine cations (along with stable counterparts) are ether excreted by kidneys or trapped by thyroid, initiating the process of thyroid hormone production. This process takes several days during which the beta radiation from I-131 is absorbed in the gland. The goal is to partly destroy the thyroid tissue, ideally so that the remaining parts ensure normal production of thyroid hormones. However, the treatment is also considered relatively successful in case of over-destruction and consequent low hormone production (the patient then regularly takes the synthetic hormone supplements). The student is not expected to perform the measurements on a patient, but to study this material and be able to follow the instructor. Basics The chapter Dosimetry/Biophysics lectures- diagnostics (accessible at the Faculty web pages) can serve as an introduction to this exercise. This section

scrutinizes the most important aspects from that material: In traversing matter the ionizing radiation exerts effects that may be only temporary (in case of nonliving material) or chronic and serious ones (living tissue). The radiation effects depend on: (i) intensity of radiation at the point of interest (ii) interaction probabilities of radiation with matter, and (iii) duration of exposure These effects are quantified by: Absorbed dose, which is the energy the radiation deposited to unit absorber mass. The unit is Grey (Gy). One Grey is 1 Joule/kilogram (J/kg) Absorbed dose indicate cumulative effects of ionizing radiation on the unit mass of the exposed medium during non-specified time spent in radiation field. Source: http://www.doksinet Dividing with time, one gets the time independent quantity; the absorbed dose rate, measured in Gy/s. One should realize that the small amounts of this energy can do a lot of harm to living tissue. This is explained by the mechanism of energy

absorption. Unlike non-ionizing radiations, which give-off energy by increasing the rate of molecular rotations and vibrations, high energy photons or particles ionize atoms, creating positive ions and free electrons. Subsequently, the ejected electrons are slowed down and recaptured by positive ions. When the radiation exposure ceases, the non-living material is usually left unchanged. However, the living tissue is sensitive to periods when free ions exist. Molecules of a living tissue continuously interact in a precise manners and ionization may interfere with these delicate processes. Radiation harms cells by ionizing biologically important molecules like DNA, which is the direct effect, or by chemically altering intracellular water, which is indirect effect. Biological damage inflicted by ionizing radiation does not depend only on absorbed dose; the type of radiation matters also. Namely, large number of ionizations within a single cell, or single macromolecule, is more harmful

than the same number of ionizations (and thus absorbed energy) distributed over more cells (molecules). Neutrons and alpha particles ionize very densely, producing more commonly multiple, irreparable damages, than gamma or beta radiation. To account for this, one defines the equivalent dose: Equivalent dose = Q.Absorbed dose where Q is quality factor. In case of alpha particles Q is 20, for neutrons range 5 to 20 applies (depending on energy), while beta particles and gamma photons have Q=1. The unit of equivalent dose is Sievert (Sv). The old unit is rem (1 rem=001 Sv) Besides, the sensitivity to radiation (radiosensitivity) depends on type of tissue (rate of proliferation is essential!). So, we also introduce the quantity called effective equivalent dose To each part of a body one assigns the specific weight factor, which accounts for differential radiosensitivity. The effective equivalent dose equals the product of the weight factor and equivalent dose in the part of body

considered. The unity is also Sievert To summarize, the effects of radiation on a living tissue depend on: Source: http://www.doksinet (i) part of body exposed (cells that divide intensively are most sensitive) (ii) absorbed dose (iii) type of ionizing radiation (heavier particles ionize more densely) (iv) absorbed dose rate (longer time enables recuperation) The examples of radiation effects are: skin burns, nausea, hair loss, sterility, cataract, bone marrow damage, changes in genes and induction of carcinoma. Excessive doses may cause death in a couple of days. However, we may also purposely expose the parts of a body to ionizing radiation in order to destroy malignant or hyper-functioning tissue. Therapy with emitters of ionizing radiation In cases of some malignant and non-malignant diseases the emitters of ionizing radiation are entered in body of a patient with intention to destroy or permanently damage the specific tissue cells. This can be achieved by placing the solid

source near the target tissue (brachytherapy), or by entering the emitter in blood, bound to specific carrier, which provides for accumulation in the target organ. In the later case, the compounds containing beta emitters are either given orally or injected intravenously. Occasionally, the carrier is the radioisotope atom (ion) itself, as is the case of 131I cation when thyroid is the target organ. Such procedures must be carefully planned in order that the damage is restricted to target tissue and to avoid both under-destruction and overdestruction (especially in case of non-malignant diseases). To accomplish this one must take into account the individual patient characteristics. Clinical problem Hyperthyroidism is a condition characterized by excessive thyroid hormone production. The surplus of thyroid hormones in body cells accelerates the metabolism in virtually all tissues. The symptoms include increased heart rate, tremor, sweating, insomnia, chronic fatigue and loss of weight,

despite increased food intake. There are several underlining diseases: diffuse toxic goitre (Graves-Basedow disease), toxic adenoma (thyroid nodule which lost pituitary control and functions autonomously), toxic multinodular goitre or, rarely, pituitary adenoma. All these diseases cause thyroid overfunctioning (i.e hyperthyroidism) and consequent increased serum concentrations of thyroid hormones (thyrotoxicosis). However, thyrotoxicosis may also be related to several other conditions/diseases. For example, some inflammatory thyroid diseases damage the Source: http://www.doksinet thyroid tissue, causing leakage of internal hormone reserves and transient thyrotoxicosis. Next, the patient with hypothyroidism (the state of insufficient production of thyroid hormones) may be overtreated with synthetic thyroid hormones, or, rarely, there may be the case of hormonally active tumours. First one establishes the diagnosis of hyperthyroidism, which is related to disorders of thyroid tissue

(i.e excludes other causes of thyrotoxicosis and abnormal pituitary functioning) The treatment than usually begins (in Europe) with antithyroid drugs. This, however, is successful only in fraction of patients. In patients who do not respond adequately to antithyroid drugs there are two kinds of definitive treatments: surgery or radioiodine. In USA, Japan, and in some European centres relatively large, so called ablative radioactive doses are planned, with goal to quickly achieve the cessation (remission) of hyperthyroid state. The price paid is thyroid overdestruction and the patient usually becomes hypothyroid. In majority of European centres one tries the individual dosimetric approach, aiming not only for remission of hyperthyroidism, but for normal thyroid function, e.g euthyroidism. Therapy planning In Croatia about 90% of hyperthyroidism is due to diffuse toxic goitre, and we will focus on this disease. The thyroid absorbed dose necessary to achieve the euthyroid state varies

from patient to patient. The reasons behind this individual radiosensitivity of thyroid tissue are not yet well understood and the success of therapy cannot be accurately predicted. Sometimes the absorbed dose is too small and the patient remains hyperthyroid (requiring retreatment), or it is too big and the patient becomes hypothyroid. Current data indicate that 120 Gy is the thyroid absorbed dose (due to beta/gamma radiation from I-131) that most frequently results in euthyroidism (in about 2/3 cases). Whether better results can be achieved by recognizing the signs of individual patient radiosensitivity is currently being investigated in several centres, including the Department of Nuclear Medicine, University Hospital Split. Since I-131 is the beta/gamma emitter the absorbed dose of 120 Gy corresponds to equivalent dose of 120 Sv. Source: http://www.doksinet For each patient one should determine the activity (radioactivity) of I-131 that, that upon oral application, ends-up in

thyroid in adequate quantity to produce the desired (planned) absorbed dose. Given the planned absorbed dose (usually around 120 Gy), the required activity increases with thyroid mass (recall that absorbed dose relates to unit absorber mass) and decreases with fraction of given activity that ends-up in thyroid (the rest is eliminated by kidneys). This fraction is measured at 24 hours after oral application of radioiodine and is called 24 h uptake. Also, the longer radioiodine spends in thyroid (the greater effective half-life of radioiodine attached to thyroglobulin molecule), the greater absorbed dose and the less radioiodine activity is required. The mass of thyroid (mthy) is assessed from ultrasonic B-mode measurements (Figure 1), of thyroid volume (Vthy), assuming that the density of thyroid tissue is 1.1 g/cm3 Thus: mthy (g) = Vthy (cm3) x 1.1 (g/cm3) To assess thyroid volume the thyroid lobes are modelled as ellipsoids, while isthmus is considered rectangular (cuboid). One

depicts the maximal longitudinal and transversal sections of thyroid, which display the maximal dimensions of thyroid lobes and isthmus. Source: http://www.doksinet Figure 1. Thyroid echograms: partial longitudinal section (upper panel) and transversal section (lower panel) Under above assumptions: Volume of each thyroid lobe = length x width x height x π/6 Volume of isthmus = length x width x height Vthy = Volume of right lobe + Volume of left lobe + Volume of isthmus Source: http://www.doksinet The measurement of 24 h uptake is performed separately using the open collimator counter. Using the same geometry, the count rate of radiation emanating from the patient neck at 24 hour after application of radioiodine are divided by the count rate due to bottle with radioactivity to be given to the patient the day before (see Exercise 4). This ratio, corrected for I-131 decay and slightly different efficacies of two measurements, is 24 h uptake (around 20% in normal thyroid, from

6080% in hyperthyroidism). These measurements can be repeated at several consecutive days From these data the effective half-life of radioiodine in thyroid is calculated. So called Marinelli’s formula relates the activity of I-131 that is required to induce the planned absorbed dose, taking into account 24 h uptake and effective half-life of radioiodine in thyroid (Appendix): I  131 activity (MBq)  planned absorbed dose (Gy) x m thy(g) x 24.7 effective half  life (days) x 24 h uptake(%) The simpler method uses the other formula: I-131 activity (µCi) = m thy (g) x 100  300 Ci 24 h uptake To apply the above formula a physician needs to choose the planned absorbed dose (for example 120 Gy). This then determines the factor 100-300 µCi in the nominator Note that the method is only semiquantitaitve one, sine the formula does not contain the effective half-life of radioiodine in thyroid. In this way one neglects the individual variations in this quantity, assuming the

empirical average value. The price paid is loss of precision; the gain is in simplicity, by avoiding multiple measurements of thyroid uptake in several days. Source: http://www.doksinet Appendix: Derivation of Marinelli’s formula The great majority of I-131 radiation energy absorbed in thyroid comes from beta electrons. So, we will first neglect the fraction of absorbed dose coming from accompanying gamma radiation. Assuming that all beta electrons are absorbed within thyroid tissue, the energy deposited equals the total number of I-131 disintegrations in thyroid multiplied by average beta electron energy (190 keV): Energy absorbed by whole thyroid = Total number of I-131 disintegrations x 190 keV We next assume that, upon oral application, all radioiodine is instantaneously trapped in the gland. Then the total number of I-131 disintegrations in thyroid equals the product of the initial I-131 activity present in thyroid (the activity orally given multiplied by thyroid uptake),

multiplied by average time the radioiodine spends in the gland (Taverage): Energy absorbed by whole thyroid = I-131 activity x 24 h uptake x Taverage x 190 keV Taverage is obtained as the weighted average of all times (from zero to infinity); the weight factor is proportional to e  ef t , where λef is the constant of effective elimination of radiotracer from thyroid (both due to physical decay and biological elimination; i.e secretion of hormones containing radioactive iodine). The constant of proportionality is obtained by normalizing the area under the curve e  ef t (t>0) to unity. By using the methods of integral calculus one obtains: Taverage = 1  ef Source: http://www.doksinet This can be written as (see Dosimetry/Biophysics lectures- diagnostics): Taverage = effective half  life ln 2 Next, by observing that the planned absorbed dose relates to radiation energy absorbed by unit thyroid mass: planned absorbed dose (Gy) = (1/ln2) x I-131 activity

(Bq) x 24 h uptake x x effective half-life (s) x 190 keV By rearranging and using more convenient units one obtains: I  131 activity (MBq)  planned absorbed dose (Gy) x m thy (g) x 26.38 effective half  life (days) x 24 h uptake(%) Finally, by taking into account that, in average thyroid, about 6% of I-131 absorbed dose is due to gamma radiation, the final Marinelli’s formula is: I  131 activity (MBq)  planned absorbed dose (Gy) x m thy (g) x 24.7 effective half  life (days) x 24 h uptake(%) There are also more accurate methods which take into account that the thyroid radioiodine uptake is not instantaneous, but the gradual one. Source: http://www.doksinet Exercise 3: Measurements in nuclear medicine Nuclear medicine utilizes many measurements and quantification of various data collected by gamma counters, single probe counting systems and gamma cameras. Using this equipment and techniques we can specify important clinical parameters, including

erythrocyte volume, plasma volume, blood volume, glomerular filtration and effective renal plasma flow, residual urine, volume of refluxed urine in children with vesicoureteral reflux. This exercise presents you with two types of gamma counters: single probe counting system and well counter. The single probe counting system (scintillation probe) consists from detector- sodium iodide crystal, a collimator and associated electronics (Figure 1). The measurements are simply done: a patient receives some activity of a radionuclide, previously calibrated by using a single probe counting system or dose calibrator. At some time after introduction, depending on the dynamics of a radionuclide, the measurements are done to register accumulation or transit of radiotracer through the organ of interest. Residual urine Residual urine is the urine that remained in bladder after urination. Under physiological conditions it should not be present. The disorder occurs mostly in elderly men, after age of

fifty, when prostate grows and squeezes urethra. The disorder is called benign prostatic hyperplasia, and symptoms are strangury, prolonged urination and chronic urinary infections. Paradoxal ishuria is an extremely difficult condition, when the bladder is full but a patient cannot urinate; the urine is leaving the bladder drop by drop and without control. Measurement of residual urine is a simple nuclear medicine procedure, using the radionuclide which is secreted by the kidneys (I-131 hipuran or Tc -99m DTPA). Radionuclide is given intravenously. Afterwards the patient drinks a lot of water (10 ml/kg) and must not urinate during the next two hours. At that time the radionuclide is eliminated from blood by renal secretion in the bladder. We register bladder activity by putting a single probe counting system over the full bladder during one minute. After this first measurement the patient urinates in the special measuring beaker, and we put again the single probe counting system over

the ˝empty˝ bladder. The difference between counts of the full and empty bladder represents voided counts. By dividing these counts Source: http://www.doksinet with milliliters of voided urine one obtains the counts per one milliliter voided. Finally, the volume of residual urine is obtained by dividing counts from ˝empty˝ bladder and counts per one milliliter voided. Renogram Renogram is time–activity curve that provides graphic representation of the uptake and excretion of a radionuclide by the kidneys. As in case of residual urine measurement, the radionuclide must be eliminated by kidneys and the single probe counting system is placed over the kidney (lumbar region). However, in contrast to residual urine measurement, the renal activity is counted immediately after injection, during the following twenty minutes, in sequences each lasting 1 minute or so. The renogram curve can be obtained for each kidney separately The renogram curve contains three phases. The first one is

the vascular transit phase. It is due to initial arrival of radionuclide in each kidney, shown as rapidly ascending curve beginning, usually lasting 30-60 seconds. The second one is the concentration phase. It represents the initial parenchymal transit of a substance, and is displayed as a decline in slope (still ascending) of the renogram, which then reaches its maximal value. Afterwards the curve declines, representing the third, excretion renogram phase (Figure 2). Depression of the second renogram phase speaks for functionally damaged kidney. Flat or even ascending third renogram phase corresponds to elimination defects (obstructive disorders) (Figure 3). Accumulation of iodine-131 in thyroid The measurement is performed similarly as in residual urine measurement, however, the single probe counting system is put over the lower front region of the neck (over thyroid gland). Thyroid gland produces thyroid hormones by using iodides for hormone synthesis, so we are using iodine 131 as

a radionuclide in this case. The iodine uptake test is performed 4, 24 and 48 hours after receiving the calibrated dose of I-131 perorally. The normal range of thyroid uptake after 24 hours is from 20 to 50 percent Hypothyroidism is a condition caused by low production of thyroid hormones in thyroid gland. Source: http://www.doksinet Lack of hormones leads to slower metabolism, the symptoms are weakness, cold, dry and itchy skin, weight gain, water retention, cold intolerance. In this case, the value of thyroid uptake is low, usually below 20%. Hyperthyroidism is the opposite condition, when thyroid gland produces excessive amounts of thyroid hormones (and needs more iodine), causing excessive concentration of these hormones in blood (and in tissues), which is called thyrotoxicosis. The value of radioiodine uptake is high, usually over 50%. The symptoms include nervousness, palpitations, tachycardia, hyperactivity, heat intolerance, tremor, sweating, weight loss, etc. The patient

receives a calibrated activity of iodine-131 by mouth. After 4, 25 and 48 hours the measurements are made by placing the single probe counting system over the thyroid and counting impulses during one minute. These numbers are then expressed as percentages of a given activity. Thus radioiodine uptake test helps in differential diagnosis of euthyroidism, hypothyroidism and hyperthyroidism. The iodine uptake test is frequently used to differentiate between two forms of thyrotoxicosis, thyrotoxicosis caused by hyperthyroidism and thyrotoxicosis not caused by hyperthyroidism. Thyrotoxicosis without hyperthyroidism is most frequently caused by inflammatory process of the thyroid gland, leading to tissue destruction and release of a large amount of thyroid hormones into the blood. This form of thyrotoxicosis lasts several weeks and is termed transient thyrotoxicosis Thus very different underlying diseases may cause thyrotoxicosis and thus produce similar symptoms. In contrast to

hyperthyroidism, thy thyroid iodine uptake in transient thyrotoxicosis is extremely low, usually around 1%. The clinical importance of differentiating these two conditions is in therapy- the thyrotoxicosis caused by destroyed tissue is only transient, requiring only symptomatic therapy, while the thyotoxicosis due to hyperthyroidism must be treated with thyrostatic antithyroid drugs. The third procedure is measurement of the effective half- life of I-131 in thyroid gland. This is done by prolonged measurements of radioiodine uptake (during several consecutive days). This is an important parameter in assessing the thyroid absorbed radiation of I-131 in hyperthyroid patients treated with radioiodine. Given the activity of I-131 orally applied, the thyroid absorbed dose increases with iodine uptake and its effective half-life in thyroid, but decreases with thyroid mass. Since the value of this parameter substantially varies from patient to patient (3-7 days), it should be included in

assessing the thyroid absorbed dose due to I-131 radiation. Source: http://www.doksinet WELL COUNTER Well counter is a heavily shielded scintillation crystal used to measure the small amounts of radioactivity contained in the small volumes (samples) of blood or urine (Figure 4). It consists of a solid cylindrical sodium iodide crystal with the cylindrical well cut in crystal, into which one places the test tube. The difference between single probe counting system and this counter is in geometry of measurement. The first one (as previously described) can register only a small portion of radiation emanating from the source (usually less tan 1%), the lower the greater the distance from the source. On the contrary, the well counter registers activity all around the source, except for the tiny fraction that escapes through the top side. The well counter is used to measure activity in blood samples in patients sent for estimation of renal function, i.e, glomerular filtration and

effective renal plasma flow At some time after intravenous application of a previously calibrated activity, the blood sample is obtained and the activity of radionuclide is measured in the well counter. The better the renal function, the lower blood concentration of a substance and thus the lower count rate of attached radionuclide is measured. Source: http://www.doksinet Exercise 4: Radiogram’s contrasts Two independent factors determine a radiogram: 1. Dimension and the type of tissue that is projected on film 2. Quality of an X-ray beam Imaging contrasts are determined by both listed factors, while an X-ray beam can be influenced by changing of the tube voltage. During recording of a radiogram the dimension and type of tissue cannot be influenced but relative frequency of photo- and Compton effects can be changed by changing the tube voltage. As a consequence, imaging contrasts also change. Instructions for exercise: 1.  Make the phantom of available parts. The phantom will

be observed using the x-ray beam  Place the phantom on the fluoroscopic device and observe the structures from outside.  Decrease voltage of the device to minimum. What do you see on the monitor? Why?  Increase voltage of the device to maximum. What do you see on the monitor? Why?  Choose optimal voltage of the device using the automatic switch. What do you see on the monitor in this case? Which structures are shown darker, and which lighter on the fluoroscopic monitor? What effect can be achieved if the tube voltage changes? Source: http://www.doksinet  Can you estimate the thickness of individual structures in the phantom if you change the tube voltage? 2.  Record the image of the phantom on the film. Which structures are shown darker, and which lighter on the Röntgens film?  What is the difference between these two methods?  What is the difference in resolution of the fluoroscopic monitor compared to the film? Why?  What are the

advantages and disadvantages when we compare these two methods with each other? Source: http://www.doksinet Exercise 5: Diagnostic ultrasound Echogram is a recording of ultrasound echo patterns that happen on the boundary of two media with different acoustic impedances (product of the media density and the rate of the sound movement in the same media). Resolution of the ultrasound display is the ability to distinguish two close details (the smallest distance at which two adjacent objects can be distinguished as separate) There are two parameters of ultrasound display: axial and lateral resolution. Axial resolution is the spatial resolution in the ultrasound beam direction. Lateral resolution is the spatial resolution along a direction perpendicular to the ultrasound beam (perpendicular to the axis of a transducer). The golas: 1. Show the basic characteristics of an ultrasound image: hypoechogen, hyperechogen, and anechogen patterns, and dynamic character of an ultrasound image

(examination). 2. Describe artifacts of an ultrasound image 3. Resolution of an ultrasound image: types of resolution and factors that influence resolution Source: http://www.doksinet Instructions for exercise:  Immerse two sticks, made of wood, in improvised water phantom (why water?). Sticks are connected on the one side and the distance between them on the other side is 1 cm (shown on the figure below).  Immerse probe in the water above sticks on the side where distance between sticks is 1 cm and detect their image in the monitor.  Move probe towards side where the distance between sticks decreases (movement should be very slow) and determine the spot where it is not possible to distinguish two sticks in the monitor anymore.  Take sticks out of the water and measure distance between them at the determined spot. That distance is called axial resolution.  Immerse sticks in water at the different depth and repeat the measurement. Does axial resolution change

depend on depth? Source: http://www.doksinet  Rotate sticks for a 90º and immerse it again in the water at the same depth like at the beginning. That will allow you to measure lateral resolution  Repeat the procedure described for axial resolution again. How does image of sticks look like in the monitor at the beginning?  When you determine the spot where it is not possible to distinguish images of two sticks in the monitor anymore take the sticks out of the water and measure distance between them at the determined spot. That distance is called lateral resolution What is the difference between axial and lateral resolution? What causes that difference?  Immerse a metal ball in the water, respectively, and find its image in the monitor. Describe the image. What artifact do you see?  Repeat the same with a prepared glove. Source: http://www.doksinet Exercise 6: Neural signal Hodgkin Huxley model of propagation of action potentials In 1952, Hodgkin and Huxley

published seminal scientific paper that describes initiation and propagation of action potential as a function of ionic conductivity of membrane channels in the nerve cells. Their contribution to the understanding of neural signal propagation is awarded with the Nobel Prize in physiology and medicine in 1964. Hodgkin-Huxley (HH) model is based on a simple model of the equivalent circuit for the part of the nerve cells: In short, the patch of membrane is modeled as the circuit consisting of a capacitor (Cm) connected in parallel with three series circuits of the battery (E) and the resistance representing ionic conductivity (g). The battery (E) is in fact the Nernst potential for a particular ion, and dynamics of ion channels in the membrane is determined by its conductivity (g). As the contributions of sodium and potassium ions in the creation of neural signals are the largest, the HH model is simplified with only three components, as shown in the figure above: sodium ion current

(INa), potassium ion current (IK), and leakage current (IL). In this way, the total current which is the sum of the capacitive current that charges the capacitor membrane and ionic currents through serial connections, ie: where dVm / dt is the change of membrane potential in time. Using ingenious trick of clamping the voltage (voltage clamp), Hodgkin and Huxley were able to Source: http://www.doksinet separate the ionic currents from the capacitor’s current in order to determine the contribution of ionic currents. This was done in the squid axon which is quite large and therefore relatively easy to experiment with. In the next step, using voltage and current clamp, H&H were able to separate ionic currents of each ion, showing them that the ionic currents of sodium and potassium follow the Ohms law (I = V / R). Thus, the ionic current of each membrane channel is proportional to its conductivity (i.e, inversely proportional to the resistance) and to the potential difference for

the ionic channel, i.e: As the total sum of the individual membrane ion currents (sodium and potassium), together with the leakage current, i.e: We obtain the following H&H model for the total current passing through the patch of the membrane: Source: http://www.doksinet Questions for the exercise: Creating an action potential as a function of ionic channel conductance Program to simulate the action potential can be found at: http://www.afodornet/HHModelhtm (turn on Java application). Prepare your desktop on your PC as follows: Tasks: 1) Determine the membrane potential of cell X which has an equilibrium potential of sodium ions of 75mV, and potassium ions -90 mV ("Reversal potential" in the "Model Options"). 2) Determine the threshold for triggering an action potential for the upper case of the cell X. Initiation occurs when the incoming (sodium) current overpass the outgoing (potassium) current. 3) Determine the refraction time for the cell X. The

absolute refractoriness is related to time of closing the sodium channels when the conductance of sodium channels drops to zero. 4) What happens with action potential when density of sodium channels is reduced by half? 5) Describe similarities and differences of the action potentials if we simultaneously: a) double the Na channel density , and reduce the density of K channels by half; b) reduce the Na channel density by half, and double K channel density; c) double both densities of Na and K channels; d) reduce both densities of Na and K channels by half. Source: http://www.doksinet Exercise 7: Audiometry - hearing test Audiometry is a procedure to determine ability to hear sounds („audire“ - listen / hear and „metria“ - measurements). It is based on a comparison with a standardized hearing thresholds obtained on a large number of people with normal hearing. Audiometric tests may not only indicate potential hearing loss, but also reveal more about the nature of hearing

loss. The main principle of audiometry lies in the measurements of responses in participants who are stimulated with certain acoustic stimuli with pre-determined parameters (eg. intensity & frequency) Audiometry based on the behavioral responses of the respondents is called subjective audiometry, while in the objective audiometry the hearing status is established by objective physiological measures such as otoacoustic emissions or evoked potentials of cortical and subcortical centers. Depending on the type of stimuli, there can be tone or speech audiometry. If hearing is tested at the limit of detection of sound, we are talking about liminar audiometry, while supra-liminar audiometry is obtained by testing at higher intensities. The result obtained from audiometry is called audiogram Liminal (threshold) tone audiometry is the basic method for testing of hearing, and it is used in this exercise. Threshold of audibility, loudness and pitch There is a minimal sound pressure that can

still be heard which is called threshold of audibility. This threshold depends strongly on the frequency. Also there is some loudest sound that someone can still bear which is called the perception limit or maximum tolerable sound pressure level. This threshold is dependent on the frequency as well. At the lowest and highest frequency that can still be heard, these two thresholds are merged. Surface in the plane of the frequency / sound pressure delimited by these two thresholds is called the auditory field. Source: http://www.doksinet Figure 1: Hearing area Figure 1 schematically shows hearing area (auditory field) and approximate area occupied by the speech conversation. Sound intensity is shown left on a linear scale in decibels (dB), where 0 dB SPL corresponds to the sound pressure of 2 * 10-5 Pa (=20 µPa) which is the typical threshold of hearing a 3 kHz tone. Keep in mind that decibel scale, indicating sound intensity, has a logarithmic dependency on the sound pressure, i.e:

Intensity level in dB SPL = 20 log ( Sound pressure [in Pa] / (2* 10-5 Pa) ) For comparison, a pressure of 2 Pa corresponds to 100 dB SPL, and 20 Pa corresponds to 120 dB (logarithmic scale!). Frequency range in which the ear can detect sound signals is approximately between 20Hz to 20 kHz. It is an extremely wide area The difference between the threshold of hearing (the highest sensitivity being around 3kHz) and the loudest sound (about 500 Hz) is even larger. Sound levels above ~ 01-02 Pa may damage your hearing It is important to distinguish the intensity of the acoustic stimuli from the subjective level of the sound sensation (loudness). Specifically, if we precisely measure the audibility thresholds with the lines of the same loudness (i.e isophones), then auditory field has a structure shown in the following figure: Source: http://www.doksinet izofone dB SPL 120 110 110 100 100 90 90 80 80 70 70 60 60 50 50 40 40 30 30 20 20 10 10 0 -10 31.25 62.5

125 250 500 1000 frekvencija (Hz) 2000 4000 8000 Figure 2: Isophones Frequency is shown on the X-axis with a logarithmic scale, whereas the y-axis is a linear scale of the sound pressure level measured in dB SPL. Reference value for the 0 dB SPL is 20Pa This sound pressure represents the threshold of audibility at frequencies with the highest sensitivity of hearing. Threshold in the Figure 2 is shown with the dotted curve Threshold is naturally optimized to the hearing limits. Namely, if any area of hearing threshold was any better, then we would be able to hear heart beats, blood flows in the capillaries or even colliding air molecules. The unit for measuring the loudness is phon. For a frequency of 1kHz, phon unit is defined to be equal to dB SPL. For example, the loudness of 10 phons is equal to the subjective loudness of a reference tone of 1kHz presented at 10dB SPL. We can then compare the loudness levels at other frequencies having the same loudness. This procedure of

determining the loudness curves depending on the frequency yields isophones. This procedure is carried out so that we present to a subject, for a short period of time, which must be longer than 200 ms (e.g 05 s) tone of 1 kHz presented at 10 dB SPL, followed by another tone of different frequency whose loudness can be regulated by the subject. Subject is requested to adjust the loudness of the second tone to match the loudness of the primary tone. When this is repeated for different frequency, we obtain isophones Looking at isophones in Fig.2, we can notice that for weak sounds (ie at low loudness levels) loudness sensitivity at higher frequencies is significantly better than for the low frequencies with the difference of 70dB or more. Source: http://www.doksinet If the loudness increases, then the "frequency response" of isophones becomes more linear, so in the example of 100-phone isophone, such difference is less than 30 dB. This means that sound pressure does not

contribute equally at low and high frequencies. This can explain why rock concerts sound good if they are loud, and quiet recordings sound "empty" because they have essentially reduced low, and to some degree high frequencies. Therefore, the ear is very nonlinear converter In order to conduct audiometry procedure, it is necessary to provide a quiet space (e.g soundinsulated cabin or sound-proof chamber) to minimize the ambient noise from interfering with the process. Also needed is calibrated audiometric system consisting of a generator of acoustic stimuli, and high-quality headphones in order to obtain reliable measurements. Exercises for practice: 1) Calibration of audiometry system consisting of a computer, a sound card, the program for audiometry (Audiometer EsserAudio Test) and headphones (Beyerdynamic, AKG) Before the audiometry, it is necessary to calibrate the system in a way that, for each frequency, by moving the cursor indicating sound volume on the computer, we

determine the intensity of sound at the threshold of audibility. The procedure involves people with healthy hearing Once the system is calibrated, it is necessary to record the calibration data in the file. 2) Determination of the threshold of audibility for frequencies from 125 Hz - 8 kHz After calibration in the first task, we are able to create audiograms. One student has the role of the examiner, and the other one the role of subject/patient. Record the results in a file What are the main features of obtained audiograms? How would you inform the patient should the obtained results indicate any degree of hearing loss? 3) Audiometry of the blocking ear. After completing the test/task in 2), the students switch roles, but in a way that a new participant carefully places earplug (or some other suitable material such as soft, but dense material such as cotton) in the test ear. Provide audiogram Discuss the results: eg if hearing loss is observed, and if so which frequencies display the

greatest loss. Source: http://www.doksinet Exercise 8: Optical bench Converging lenses Lenses are transparent optical objects confined by two spherical surfaces. Optical axis is the straight line passing through the geometrical centre of a lens and joining the two centers of curvature of its surfaces. That is axis of rotational symmetry Two characteristic points for each lens are located at the optical axis. They are called focal points or focuses If a lens is thin and located in air, distances of both focuses from the lens center are equal. They are called focal lengths (Figs 1 and 2). If an object is located at an infinite distance its image will be formed at the focal point F 2 of the lens (Fig. 1) Figure 1. If an object is located at the focal point F1 of the lens its image will be formed at the infinite distance (Fig. 2) Figure 2. Source: http://www.doksinet The distance between the converging lens and the image, nature and the image size depend on the distance between

the converging lens and the object. The formation of image by converging lens is presented in Fig. 3 Divergent ray beam spreads from the point A located at the top of the real object . Refraction of three characteristic rays by the converging lens is described Ray 1 runs parallel to the axis until it reaches the lens; then it refracts through the lens and leaves along a path that passes through the lens focus. Ray 2 runs straight through the center of the lens never bending. Ray 3 first passes through the focus until it reaches the lens; then it refract through the lens and leaves parallel to the lens axis on the other side of the lens. After refraction all this rays will meet in point A′. Images of all points located at the object principle. Figure 3. can be formed using the same Source: http://www.doksinet If we denote the distance between the converging lens and the object distance between the converging lens and the image with x1, and the with x2, than the equation that

connects this values with focal length, f, can be defined as: (1) Equation (1) is called the lens conjugate equation in air. Distances (x1, x2 i f) in this equation are measured from the lens center (0). In order to determine their sign it is necessary to know the light beam direction (Fig. 3) Quantities that have the same direction like the light beam have positive sign. If they have opposite direction the sign is negative According to this, the distance x1 in this case will have negative, and x2 positive sign (Fig. 3) If some object is moved from infinity towards the lens, formed images for different object positions can have different position, size and nature. An overview is presented in Table 1 Table 1. x1 ∞> > 2f = 2f 2f > >f <f x2 Image f < x2 < 2f real, inverted, smaller in size x2 = 2f real, inverted, the same size as object x2 > 2f real, inverted, bigger in size x2 < 0 imaginary, upright, bigger in size The lens strength, j, in

diopters is defined as the inverse of the focal length, f, in meters if the lens is located in air: (2) Source: http://www.doksinet Example 1. Determine the focal length and the strength of converging lens Choose distance x1 at the optical bench and determine x2. Using Eq (1) and (2) calculate f and j. It is necessary to determine the object and image size for each measurement Symbols for them are y1 and y2, respectively. If the image is inverted y2 has negative sign Linear magnification, m, is defined as ratio . Eq (3) is valid if measurements are performed accurately: (3) Make four measurements for the same lens, two for image smaller in size and two for image bigger in size. Instructions for measurements with real object: All measurements will be performed on the optical bench (Figs. 4 and 5) That is metal bar with millimetric scale. The scale serves to measure parameters x1 i x2 All used holders can slide over the bar. Screen Z is a white metal board where an image forms

An arrow is cut out in a piece of thin metal and used as an object. In order to perform an experiment it has to be illuminated by the source of light. Picture of the whole optical bench is presented in Fig 4 All used components are presented in Fig. 5 schematically Source: http://www.doksinet Figure 4. Figure 5. In all measurements the lens height should be adjusted in a way that its optical axis passes through the center of the object (Fig. 5) Additionally, three planes perpendicular to the optical axis (of the lens (R), object (P) and screen (Z)) should be always parallel. It can be adjusted easily because all handlers (D) in holders (S) can be rotated, elevated or lowered. Handlers (D) can be tightened by screws (V) at the chosen positions Source: http://www.doksinet Put the converging lens Lk in the holder 1 and choose the parameter x1 arbitrarily. Then put the screen (Z) in the holder 2. Move it back and forth in order to find clear image of the object Use labels on the

bottom of each holder (S) to accurately read all positions on the scale. Measure quantities x1, x2 i y2 and determine their signs precisely. Write measured quantities in Table. 2 Calculate the focal length , linear magnification m (use two ways described in Eq. 3) and the strength, j, of the converging lens Table 2. Measurement x1/cm x2/cm y1/cm y2/cm f/cm j/dpt number (4) (5) Source: http://www.doksinet Example 2. Determine the focal length and the strength of converging lens using method of virtual object Imagine converging beam of rays falling on the converging lens (Fig. 6) These rays would be focused in one point (A) if the converging lens was absent. In the presence of the converging lens the point A represents the virtual object for that lens. Figure 6. The virtual object for the lens L can be created in manner presented in Fig. 7 The auxiliary lens L′ forms the converging beam of rays that falls on the main lens L. Rays 1, 2 and 3 would have intersected at

the point A′ if the lens L would be absent. In that case point A′ would have been the image of the real point A (this point belongs to the real object ). In the presence of the lens L point A′ becomes virtual object at the distance x1 from that lens. The image object is formed at the distance x2. Figure 7. of the virtual Source: http://www.doksinet Quantities x1 i x2 in this example, according to previously described rule, have positive signs. Instructions for measurements with virtual object: Measure the focal length, f, and strength, j, for the converging lens from previous example using method of virtual object (Figs. 8 and 9) Figure 8. Put the auxiliary lens L′ in the holder 1, screen in the holder 3 and adjust their heights. Choose arbitrarily the distance between the object P and the lens L′. Find clear image on the screen That is the virtual object for the lens L. Measure its height, y1 The lens L′ and screen Z should not be moved until you determine

quantity x1. Source: http://www.doksinet Figure 9. Choose the position of the main lens L arbitrarily between the auxiliary lens L′ and the screen Z (Figs. 10 and 11) Adjust its height The image will disappear from the screen Measure the distance x1 between the lens L and the screen Z. Figure 10. Figure 11. Source: http://www.doksinet Move the screen towards the lens L until you see clear image again (Figs. 12 and 13) Measure the distance x2, between the lens L and the image, and its height y2. Figure 12. Figure 13. Make three measurements for three different values of x1. Write all measurements in the table prepared like in the example 1. Calculate f and j Source: http://www.doksinet Example 3. Determine the focal length and the strength of converging lens using Bessel’s method If one particular distance between an object and screen (l) is adjusted, it is possible to find two lens positions, I and II, for which the image of the same object forms. Bessel’s method

is based on this fact. One time formed image is smaller and another time is bigger than the object (Fig 14) Figure 14. Position I: f < a < 2f, b > 2f the image is inverted and bigger in size, Position I: a > 2f, f < b < 2f the image is inverted and smaller in size. The distance g between both lens positions can be easily measured. From Fig 14 can be concluded that: (6) (7) Adding and subtracting Eqs. (6) and (7) gives expressions for quantities a and b: (8) (9) Source: http://www.doksinet Combining Eqs. (8) i (9) with Eq (1) give an expression for focal length: (10) Determine the focal length f of the converging lens for four different l values using Bessel’s method. Instructions for measurements: Choose the distance between the object and screen (l). It should be bigger than 4f Put the lens between them and move it towards the object until you find clear image on the screen (this image is bigger in size compared to object). Read the lens position I After

that move the lens towards the screen until you find again clear image (this time the image is smaller in size compared to object). Read the lens position II Use these two positions to calculate the distance between them, g. Measure the distance between the object and screen (l) Use Eq (10) to calculate the focal length f. Determine the lens strength j Table 3. Measurement number m Which one of this three methods should give more accurate results and why? Source: http://www.doksinet Exercise 9: Viscosity of fluids Laminar flow occurs when a fluid flows in parallel layers and parallel to side walls of a tube, with no disruption between layers (Fig. 1) Figure 1. Particular layers in real fluids do not have the same velocities because of the inner friction forces present inside of a fluid. The viscous force between layers is equal: (1) A – the cross section area of tube v – the velocity of a particular layer x – the distance of a particular layer from the axis of tube Δv/Δx

– the rate of change of velocity with the distance (the velocity gradient)  – the coefficient of viscosity Source: http://www.doksinet It is possible to determine the unit for the coefficient of viscosity from equation (1): (2) Layers of a fluid located immediately by the side wall of tube are at rest, v(R)=0, while the layer whose flow coincide with axis has maximal velocity, v(0)=vmax. Distribution of velocities of all layers is parabolic function. It depends on parameter x (the distance of a particular layer from the axis of tube): (3) Newtonian fluids are real fluids in which viscosity is independent on volume flow rate at certain temperature. Volume, V, of such fluid can be calculated according to Poiseuille’s law: (4) r – the radius of tube Δp – the pressure difference between the ends t – time  – the coefficient of viscosity l – the length of tube Source: http://www.doksinet Instructions for exercise: In this exercise we measure the relative

viscosity of unknown fluid and water. Ostwald viscometer, shown in Fig. 2, is used Ostwald viscometer is U tube made of glass and has two unequal arms Fluid has to be poured in the broader arm. Reservoir of volume V is located close to the top of the narrower arm. It is confined between labels a and b We will measure a time required for water or fluid level to descend between labels a and b; i.e time required to elapse volume V of certain fluid Figure 2. Using Eq. (4) it is possible to determine the coefficient of viscosity of unknown fluid, ηt It is necessary to use equal volumes of water and unknown fluid in all measurements. Based on this we can use following equations: (5) The pressure difference between the ends can be determined using Eq. (6): (6) Source: http://www.doksinet If we combine equations (5) and (6) the formula for the relative viscosity can be derived: (7) Or: (8) Example 1. Determine the relative viscosity of unknown fluid compared to water using Ostwald

viscometer The density of an unknown fluid is written on the bottle. Take 10 ml of water using a pipette and pour it in the broader arm of the viscometer. Elevate water level in the narrower arm above the label a. Let water to flow Turn on a stop-watch at the moment when the water level elapses by label a and turn it off when it elapses by label b. Repeat the same measurement a few times The next step is to empty the viscometer and rinse it with few milliliters of unknown fluid. Repeat the same measurements with 10 ml of an unknown fluid a few times. Write all results in table 1 Table 1. Measurement number Water tv (s) Fluid Δtv (s) Based on this data calculate the relative viscosity using Eq. (8) tt (s) Δtt (s) Source: http://www.doksinet Example 2. Compare viscosities of two sugar solutions, glucose and Soludex 40 Glucose is a primitive sugar – monosaccharide. Soludex 40 is polysaccharide composed of large number of glucose molecules. Its molecular weight is 40000

Daltons Notice that molecular weights of two described sugar solutions differ drastically. Measurements will be performed for four different concentrations of glucose and Soludex 40 (40, 20, 10, and 5%) in a way described in Example 1. Twenty percent solution means 20 g of a sugar per 100 ml of solution. Calculate both viscosities, for glucose and Soludex 40, relative to water Write all results in the table 2. and present it graphically using millimetric paper Which solution has higher viscosity? Compare the results for solutions with the same molecular weights. What can be concluded if we compare the same molar concentrations of glucose and Soludex 40? One important physiological quantity depends on molar concentration of dissolved substance. Which one? How does the relative viscosity change depend on concentration of dissolved substance? Comment the final results. Table 2. Concentration c(%) Glucose tv (s) Soludex 40 Δtv (s) tt (s) Δtt (s) Source: http://www.doksinet 6

rel 4 2 0 0 20 40 60 80 100 c (%) Graph 1. Example 3. Repeat all measurements using solution from Example 1 at different temperatures Present graphically the dependence of viscosity on temperature and comment the final result. Source: http://www.doksinet Exercise 10: Hemodynamics Goals:  Illustration of basic hemorheologic laws: (i) proportionality between the arterial pressure and blood flow and (ii) approximate proportionality between the left ventricular stroke volume and pulse pressure (difference between systolic and diastolic pressure),  demonstration of a possibility to noninvasively assess the hemorheologic parameters, instead of a complex and invasive direct measurement,  illustration of a modeling in medicine, when one uses reasonable approximations to relate the physiologic parameters with quantities that can be measured simply and accurately,  gaining an insight in practical problems of measuring the arterial pressure and heart rate at rest and

during exercise, as well as in measurement errors and ways to assess these errors. BACKGROUND This exercise shows and quantitatively assess that the total peripheral resistance (R) decreases during aerobic exercise, the more the level of exercise is. One uses the approximate proportionality of the left ventricular stroke volume (SV) and pulse pressure; a difference between systolic pressure Ps and diastolic pressure, Pd; P=Ps-Pd. The left ventricle ejects the majority of its stroke volume (about 80%) in the first third of systole, during the fast ejection phase. During this time the pressure increases from minimal (diastolic) to maximal (systolic) value, which enlarges the aorta from minimal to maximal volume, depending on its compliance (Caorta ). Within bounds of elastic proportionality the systolic increase in aortic volume is: systolic increase in aortic volume = P x Caorta (1) Source: http://www.doksinet The systolic increase in aortic volume is less than its stroke

volume because (i) the volume of blood ejected in the period of fast ejection is only 80% and (ii) some of this output leaves aorta and pushes the blood through capillaries. Since the peripheral perfusion is nearly constant, the later part, corresponding to 1/3 of a systole, which itself is 1/3 of a cycle at rest, is thus only about 1/9 of a stroke volume. Thus: systolic increase in aortic volume = SV x k (2) where, roughly, the parameter k is 0.8 x 089  07 Combining equations (1) i (2): P = SV x k/Caorta (3) Equation (3) expresses the proportionality between pulse pressure and left ventricular stroke volume (normally it is also the stroke volume of the right heart!). The proportionality constant is the part of the systolic increase in aortic volume in stroke volume, divided by aortic compliance. In other words, the greater the stroke volume the greater the pulse pressure, which increases with intensity of fast ejection and decreases with aortic compliance. Equation (3)

cannot be directly used to assess the stroke volume in a person since aortic compliance varies between people; in a given person it decreases with age and between different hemodynamic conditions (it decreases in exercise and other conditions that increase the stroke volume, when aortic collagen fibers stretch extensively). In addition, the parameter k in equation (3) also depends on hemodynamic state and decreases in physical stress. In physical stress (and other conditions of increased cardiac output) the systole occupies larger portion of heart cycle (absolute duration is fairly constant), so that larger portion of stroke volume goes to periphery during fast ejection (more than 11%), which itself accounts for less than 80% of stroke volume. However, applying equation (3) to the same person in two different hemodynamic states, the ratio of stroke volumes is: SV1/SV2 = (P1/P2)x(Caorta /k)1/(Caorta /k)2 (4) Source: http://www.doksinet Since exercise decreases both parameter

k and aortic compliance, it may be reasonable to assume that their ratio does not change much in the same person between rest and exercise. This is the basic assumption of our model, enabling us to equate the ratio (Caorte /k)1/(Caorte /k)2 in equation (4) to unity. This simplifies the equation (4) to: SV1/SV2 = P1/P2 (5) Equation (5) enables to assess the exercise induced relative changes in stroke volume from relative changes in pulse pressure. Since the cardiac output (CO) equals the product of stroke volume and heart rate (f), equation (5) also gives: CO1/CO2 = (P1/P2) x (f1/f2) (6) Equations (5) and (6) relate the quantities that require complex, invasive measurements (stroke volume, heart output) with quantities accessible simply and noninvasively (systolic and diastolic blood pressure, heart rate). Equating the right atrial pressure to zero, the stroke volume equals the mean arterial pressure (Pa) divided by total peripheral resistance (R): CO = Pa/R (7)

From relations (6) and (7) it follows that the ratio of peripheral resistances in a person in two hemodynamic states is: R1/R2 = (P2/P1)x(f2/f1)xPa1/Pa2 If indexes 1 and 2 denote stress and rest respectively, the above equation translates to: (8) Source: http://www.doksinet Rstress/Rrest = (Prest/Pstress)x(frest/fstress)xPastress/Parest (9) The common noninvasive manometers cannot measure the blood pressure continuously, but only the extreme values, the systolic and diastolic pressure. Continuous noninvasive measurement of arterial pressure is possible by the method of photopletismography. Since the method is not accessible to us, we will assess the mean arterial pressure as the weighted average between systolic and diastolic pressure. The weight factors are the parts due to systole or diastole in total duration of cardiac cycle. We can assume that at rest systole lasts 1/3, and in stress 1/2 of cardiac cycle With these assumptions: Pastress=

Pdstress+(1/2)Pstress and Parest= Pdrest+(1/3)Prest and equation (9) becomes: Rstress/Rrest = (Prest/Pstress)x(frest/fstress)x(Pdstress+(1/2)Pstress)/(Pdrest+(1/3)Prest) (10) Experiment The exercise is done in pairs; two students exchange the roles of examinee and experimenter. Basal measurements -subject is relaxed and stands still; -by sphygmomanometer the experimenter measures the systolic and diastolic pressure on the dominant arm; -by palpating the radial artery the experimenter measures the heart rate; she (he) uses the stop-watch to measure the time elapsed during 30 consecutive beets (why is this method superior to usual method of counting beats in predefined time?). Source: http://www.doksinet Stress -Degree I: 30 seconds of running at the spot, with maximal knee elevations, but relatively slow, approximately 3 ‘steps’ in a second. -Degree II: the same, but faster, about 6 "steps" in a second. Student should wear light clothing (no tight

trousers) and adequate shoes (light, flat bottom). Post exercise measurements -Measurements are performed immediately after each exercise; after completing the first part, there is a pause, at least until the heart rate returns to basal value. -In order to complete the evaluations as soon as possible the examinee himself measures his heart rate, while his college measures the arterial pressures on the opposite arm. -Unlike in basal measurement, the heart rate is measured during only 15 beats (why?). DATA ANALYSES: 1. Use the measured hemodynamic parameters and equations 5, 6 and 10 to calculate the relative changes in cardiac output, stroke volume and total peripheral resistance for each degree of exercise. 2. Use a predefined table to write down the measured and assessed parameters 3. Comment on the association between measured and assessed hemodynamic parameters in basal state and after exercises; compare the degrees of exercise. 4. Compare your results with results of the

colleagues (who exercised most vigorously, did she (he) also obtained the largest post exercise decrease in total peripheral resistance?). 5. Consider the model limitations and measurement errors; which of those errors can be assessed and how? 6. How can one decrease the measurement errors? 7. Can one improve the model of noninvasive assessment of relative changes in hemodynamic parameters (eg. by using continuous blood pressure measurements or parallel ECG recording)?